Instructions for use Oxylog 3000 plus SW 1.n 123
Reprocessing
Disassembling the breathing valve
5 To loosen the cover (A), turn the cover
counterclockwise by approx. 90°.
6 Remove the silicone diaphragm (B).
Removing the disposable breathing
circuit for adults
1 Disconnect flow measuring lines (A and B).
2 Disconnect the breathing hose (C).
3 Correctly dispose of the complete disposable
breathing circuit. Refer to the chapter
"Disposal" on page 137.
045rubberdisc
WARNING
Risk of CO2 rebreathing
The rubber disc (A) in the housing must not be
removed, damaged or bent, otherwise the
valve will not work properly and will lead to a
risk of patient injury.
A
B
A
047
CAUTION
The disposable breathing circuit must not be
cleaned, disinfected or sterilized: it cannot with-
stand high temperatures and may be damaged.
A
B
C