Instructions for use Oxylog 3000 plus SW 1.n 79
Operation
Setting the HME correction
The temperature and moisture of the heat and
moisture exchanger (HME) have an effect on the
flow measurement. Oxylog 3000 plus can
compensate for the presence of an HME.
When using HME, select, set and confirm
HME On in the Settings window with the
rotary knob.
When HME On is selected, the flow sensor
assumes an expiratory gas temperature of 35 °C
and a relative humidity of 0 %.
When HME Off is selected, the flow sensor
assumes an expiratory gas temperature of 37 °C
and a relative humidity of 100 %.