Getting started
60 Instructions for use Oxylog 3000 plus SW 1.n
If the check was not successful:
The Oxylog 3000 plus displays the alarm Filter
check failed. The measured CO
2 value is outside
the permissible tolerance range.
Check whether the sensor (A) or test filter (B) is
soiled and clean them if necessary. Repeat the
CO2 filter check.
For information on connecting the CO
2 sensor and
cuvette, see page 48. For information on CO2 zero
calibration and the filter check during ventilation,
see page 91. For information on the CO
2
measurement see page 91. For information on CO2
configuration in Customer Service Mode, see
page 103.
Preparation for use after system check, CO2 zero calibration and
CO2 filter check
1 Assemble Oxylog 3000 plus for operation.
Refer to "Assembly" on page 35.
2 Connect to the power supply and gas supply.
Refer to "Internal rechargeable battery"
on page 37 and "Connecting the gas supply"
on page 41.
3 Start the ventilator:
Select Start ventilation from the menu and
confirm.
Or
Press the Alarm Reset key.
14737171