Instructions for use Oxylog 3000 plus SW 1.n 71
Operation
– Insp.term.%PIF
Inspiration termination criteria of pressure
supported breaths as percentage of the peak
inspiratory flow (PIF).
– Hose type
The selected hose type must match the hose
type in use. Otherwise a correct volume
measurement cannot be guaranteed.
–CO
2 cuvette type (optional)
PCBIPAP/PS
The following parameters can be adjusted
on-screen in addition to the parameters of
ventilation mode PC-BIPAP:
– Setting on page 1: Pressure support
Δ
Psupp
above PEEP
– Setting on page 2: Pressure rise time Slope
Flat slope = slow pressure rise
Medium slope = medium pressure rise
Steep slope = rapid pressure rise
1403717115037171
NOTE
The cuvette windows of the reusable cuvette and
disposable cuvette have different optical prop-
erties. Therefore, the correct cuvette type must be
selected in the settings menu. Otherwise the zero
point will be shifted by as much as ±8 mmHg CO
2.
12437171