Instructions for use Oxylog 3000 plus SW 1.n 41
Assembly
Connecting the gas supply
Caution when handling O2:
Supply from an O2 cylinder
1 Connect the pressure reducer (270 to 600 kPa
outlet pressure, 500 kPa nominal pressure) to
the O2 cylinder.
WARNING
Risk of explosion
Protect the O2 cylinder against tipping over.
Keep away from excessive heat.
WARNING
Risk of fire
Do not grease or lubricate O
2 fittings, such as
cylinder valves and pressure reducers and do
not handle with greasy hands.
WARNING
Operate cylinder valves by hand and rotate
slowly to prevent the risk of fire or explosion.
Do not use tools.
WARNING
Only use medical grade oxygen.
WARNING
To prevent risk of fire, always provide
adequate ventilation in the area where the
ventilator is being operated in order to
maintain the ambient O
2 concentration
below 25 %.
WARNING
No smoking or open flames.
O
2 enhances combustion of other substances
and can intensify fires.
WARNING
Always use gas cylinders and pressure
reducers that comply with all applicable
regulations.
WARNING
Risk of suffocation
Always use filled O2 cylinders.
WARNING
To prevent damage to the ventilator due to
excessive O2 supply pressure on the input,
only use a pressure reducer with a blow-off
valve at the outlet that limits the outlet
pressure to a maximum of 1000 kPa in case of
a malfunction.