Getting started
54 Instructions for use Oxylog 3000 plus SW 1.n
Checking the operational readiness
The device check must be carried out in the
following situations:
– Before every use of the device if the breathing
hose was changed.
– At least every six months.
Oxylog 3000 plus interrupts the device check if a
fault is detected.
The relevant fault is indicated on the screen.
Performing the device check
The device check consists of the following steps:
– Connecting the test lung
– Switching the device on
– Checking the connections
– System check
– Checking the power supply failure alarm
The duration of the device check is approximately
3 minutes.
Connecting the test lung
Reusable breathing circuit for adults
WARNING
The device check must be carried out with the
ventilation accessories that will subsequently
be used for ventilation. Make no changes
afterwards, otherwise there is the risk of inac-
curate flow measurement and/or insufficient
ventilation.
WARNING
The patient may be endangered if the device
check is not completed successfully.
021