Instructions for use Oxylog 3000 plus SW 1.n 81
Operation
Shutdown
After disconnecting the patient, switch the ventilator
off.
1 To switch the device off, hold down the
key (A) for approximately 3 seconds.
Ventilation is now stopped and a high-priority
alarm is issued.
This alarm can be silenced with the key.
2 Either:
Press the rotary knob (B) to confirm switch
off.
Or
Press the key (A) to resume ventilation
with the previous settings.
3 With gas supply from an O
2 cylinder:
Close the cylinder valve.
With gas supply from the central gas supply
system:
Disconnect the high-pressure connection from
the source.
4 Disconnect the breathing hose.
00337170
NOTE
When the device is switched OFF, the battery is
still being charged if the device is connected to
an external power source.
Oxylog 3000 plus
AB
WARNING
The cylinder valve must be closed completely
to avoid gas flow leakage by the device.
048
CAUTION
When disconnecting the breathing hose from the
device, always grip the sleeve (A) and never the
corrugations (B).
If this is not done, the corrugations or hose may
be torn from the sleeve.
A
B

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals