Reprocessing
126 Instructions for use Oxylog 3000 plus SW 1.n
Prerequisites:
The washer-disinfector has been prepared in
accordance with the manufacturer's
instructions.
Positioning the components in the load carrier
Required aids:
Short breathing hose (e.g., 35 mm silicone
breathing hose, order no. 2165619) for
connection of the breathing valve
Procedure:
1 Securely position items in the basket.
2 Connect the hoses and measuring lines to
suitable nozzles of the carriage in the washer-
disinfector.
3 Connect the breathing valve (without cover and
silicone diaphragm) to a suitable nozzle of the
carriage in the washer-disinfector using a short
breathing hose.
4 Ensure the following:
All surfaces and interiors can be completely
rinsed.
The water can drain off freely.
Performing reprocessing:
1 Select a cycle.
2 When the cycle has ended, check the
components for visible soiling and repeat the
cycle if necessary.
3 Check the components for visible damage and
replace if necessary.
Storage and transport
After reprocessing, there are no special
requirements for the storage and transport of the
product. All other information in the instructions for
use regarding storage and transport continues to
apply. In addition, the requirements that prevent
contamination or damage to the product must be
met. These include, for example, dry and dust-free
storage and avoiding damage during transport to
the operating location.
Further disinfectants and
reprocessing procedures
Disinfectants
Use disinfectants that are nationally approved and
are suitable for the particular reprocessing
procedure.
Step Agent Manufacturer Concentration Temperature Contact time
Preliminary
cleaning
Tap water - - Tap water tem-
perature
Min. 2 min
Cleaning MediClean/
MediClean
Forte
Dr. Weigert In accordance
with the manu-
facturer's
instructions
Min. 55 °C
(131 °F)
Min. 10 minutes
Flushing Demineralized
water
- - Tap water tem-
perature
Min. 1 minute
Disinfecting - - - Min. 93 °C
(199.4 °F)
Min. 10 minutes
Drying - - - - Drying time
depends on the
load

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals