Alarms
84 Instructions for use Oxylog 3000 plus SW 1.n
Safety information
Position of user relative to alarm system
The alarm system is designed so that the user can
see the alarm messages from a distance of 1 m
(39 in).
The specified alarm tone volumes apply to a
distance of 1 m (39 in) at normal ambient sound
volume.
Alarm priorities
The actual screen display may differ in appearance
or configuration.
Oxylog 3000 plus assigns priorities to alarms. The
alarm message is highlighted with the
corresponding number of exclamation marks. Also,
different acoustic alarm signals are generated for
the various alarm priorities.
!!! = Warning
!! = Caution
!= Note
For information on how to remedy alarm causes,
refer to the table "Alarm – Cause – Remedy"
on page 108.
WARNING
Risk when failing to hear alarms
The alarm volume must be adapted to the sur-
roundings (see page 80), and the user must
stay within hearing range of the acoustic
alarm signal.
This allows for rapid detection of the alarm
and appropriate response.
WARNING
Pay additional attention in environments
where the surrounding noise interferes with
hearing the maximum alarm tone of the device
(e.g. in a helicopter).
036