EasyManuals Logo

Dynamic DX User Manual

Dynamic DX
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #117 background imageLoading...
Page #117 background image
MT
117
7.2 Localised parameter storage
The parameters of the DX System are stored in several locations:
The Master Remote (these parameters are also called UCM parameters)
The Power Module
Each intelligent DX Module stores its own parameters
The Master Remote stores all parameters that relate to speed, direction and user
preferences, for example:
Speed, acceleration and deceleration parameters
Joystick sensitivity parameters
Drive profiles
The Power Module stores the parameters that are needed to keep the system safe,
and to convert the speed and direction commands of the Master Remote to the
correct voltage and the correct current needed to drive the motors. For example:
Maximum temperature parameters
Parkbrake test and motor test parameters
Load Compensation,
compensates for a powerful or a weak motor
Veer Compensation,
compensates if the two motors used are not behaving exactly the same.
Intelligent DX Modules store their own parameters, for example:
The 5-switch and the Sip and Puff secondary remotes
The IRIS infrared transmitter

Table of Contents

Other manuals for Dynamic DX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dynamic DX and is the answer not in the manual?

Dynamic DX Specifications

General IconGeneral
BrandDynamic
ModelDX
CategoryNetwork Router
LanguageEnglish

Related product manuals