• Support added for turning the local display either on or off from the local display
• Corrected the velocity calculation to use actual density rather than standard gas density when the
transmitter is in Gas Standard Volume mode
• Support added so users can reset individual totalizers from the local display
• Corrected the water density values in the standard Plato curve for concentration measurement
• Support added to the local display so alert conditions can be set to detect (enabled)
• Enhanced HART interface so HART bursting no longer stops when HART commands are received for
addresses other than the current address
• Enhanced local display to show accurate readings when Frequency Output mode is set to dual pulse with a
0° phase shift
• Improved concentration measurement curves so temperature values are unit converted when reading the
data through HART
Feature additions:
• ProLink III v4.0 contains support for the applicable additional features noted below. If you have ProLink III
v3.0 installed, go to the ProLink Product page by accessing www.emersonprocess.com/micromotion/
softwaredownloads.
• Added support for the following RS-485 printers:
— Epson
®
TM-T88
— Epson TM-U295
— Digitec 6610A
— Generic
— Terminal
• Added the following HART support (contained in 5700 HART DD for device revision 2 and 5700 HART DTM
for device revision 2):
— Added support to HART master to poll for standard gas density and water cut
— Implemented HART APM commands
• Added support for the following features:
— 2-stage batcher
— Totalizer history
— Enhanced SMV functionality
— Changed SMV aborted alert so it clears when alert is acknowledged
— Changed APM drive gain threshold determination so it does not run when SMV is running
— Added NAMUR NE 53 revision to local display
— Added a screen to the local display to warn the user to put their loop into manual before making
configuration changes
— Implemented Modbus function 43 or FDI identification functionality
— Created ETO 27577 to disable Gas Density Correction for Large Meters