Commissioning Proline Promass E 500 PROFIBUS PA
126 Endress+Hauser
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Maximum frequency value The Frequency option is
selected in the Operating
mode parameter (→ 121)
and a process variable is
selected in the Assign
frequency output parameter
(→ 125).
Enter maximum frequency. 0.0 to 10 000.0 Hz –
Measuring value at minimum
frequency
The Frequency option is
selected in the Operating
mode parameter (→ 121)
and a process variable is
selected in the Assign
frequency output parameter
(→ 125).
Enter measured value for
minmum frequency.
Signed floating-point
number
Depends on country
and nominal
diameter
Measuring value at maximum
frequency
The Frequency option is
selected in the Operating
mode parameter (→ 121)
and a process variable is
selected in the Assign
frequency output parameter
(→ 125).
Enter measured value for
maximum frequency.
Signed floating-point
number
Depends on country
and nominal
diameter
Failure mode The Frequency option is
selected in the Operating
mode parameter (→ 121)
and a process variable is
selected in the Assign
frequency output parameter
(→ 125).
Define output behavior in
alarm condition.
• Actual value
• Defined value
• 0 Hz
–
Failure frequency In the Operating mode
parameter (→ 121), the
Frequency option is selected,
in the Assign frequency
output parameter (→ 125)
a process variable is selected,
and in the Failure mode
parameter, the Defined value
option is selected.
Enter frequency output value
in alarm condition.
0.0 to 12 500.0 Hz –
Invert output signal – Invert the output signal. • No
• Yes
–
* Visibility depends on order options or device settings