Proline Promass E 500 PROFIBUS PA Commissioning
Endress+Hauser 131
Parameter Prerequisite Description User interface /
Selection / User
entry
Factory setting
Assign limit The Limit option is selected in
Relay output function
parameter.
Select process variable for limit
function.
• Mass flow
• Volume flow
• Corrected volume
flow
*
• Target mass flow
*
• Carrier mass flow
*
• Target volume
flow
*
• Carrier volume
flow
*
• Target corrected
volume flow
*
• Carrier corrected
volume flow
*
• Density
• Reference density
*
• Reference density
alternative
*
• GSV flow
*
• GSV flow
alternative
*
• NSV flow
*
• NSV flow
alternative
*
• S&W volume flow
*
• Water cut
*
• Oil density
*
• Water density
*
• Oil mass flow
*
• Water mass flow
*
• Oil volume flow
*
• Water volume
flow
*
• Oil corrected
volume flow
*
• Water corrected
volume flow
*
• Concentration
*
• Temperature
• Oscillation
damping
• Pressure
• Totalizer 1
• Totalizer 2
• Totalizer 3
–
Assign diagnostic behavior In the Relay output function
parameter, the Diagnostic
behavior option is selected.
Select diagnostic behavior for
switch output.
• Alarm
• Alarm or warning
• Warning
–
Assign status In the Relay output function
parameter, the Digital Output
option is selected.
Select device status for switch
output.
• Partially filled pipe
detection
• Low flow cut off
• Digital output 4
*
–
Switch-off value The Limit option is selected in
the Relay output function
parameter.
Enter measured value for the
switch-off point.
Signed floating-point
number
Depends on country:
• 0 kg/h
• 0 lb/min
Switch-off delay In the Relay output function
parameter, the Limit option is
selected.
Define delay for the switch-off
of status output.
0.0 to 100.0 s –
Switch-on value The Limit option is selected in
the Relay output function
parameter.
Enter measured value for the
switch-on point.
Signed floating-point
number
Depends on country:
• 0 kg/h
• 0 lb/min