Commissioning Proline Promass E 500 PROFIBUS PA
130 Endress+Hauser
Relay output function
→  130
Assign flow direction check
→  130
Assign limit
→  131
Assign diagnostic behavior
→  131
Assign status
→  131
Switch-off value
→  131
Switch-off delay
→  131
Switch-on value
→  131
Switch-on delay
→  132
Failure mode
→  132
Switch status
→  132
Powerless relay status
→  132
Parameter overview with brief description
Parameter Prerequisite Description User interface / 
Selection / User
entry
Factory setting
Terminal number Shows the terminal numbers
used by the relay output
module.
Not used
24-25 (I/O 2)
Relay output function Select the function for the
relay output.
Closed
Open
Diagnostic
behavior
Limit
Flow direction
check
Digital Output
Assign flow direction check The Flow direction check
option is selected in the Relay
output function parameter.
Select process variable for flow
direction monitoring.