FT-111, User Manual Rev. 2.0.0, June 2022 Page 85 of 137
L
Read the all data logging in the SD card
Command : [ADR][L]
Response : [ADR][L][ ALL DATA]
Example :
Command : 01L
Response : 01LA[ALL DATA] (All data logging is sending) or
01LN (SD card could not read or there is no SD card)
Comments: All data logging sends after acknowledging (ACK).
N
Read the scale name
Command : [ADR][N]
Response : [ADR][N][STATUS][ 16 digits ]
Example :
Command : 01N
Response : 01NASCALE-1 (Scale name is SCALE-1)
Comment: Scale name is transmitted immediately after receiving command.
G
Read voltage value of DC power supply
Command : [ADR][G]
Response : [ADR][G][STATUS][VOLTAGE VALUE]
Example :
Command : 01G
Response : 01GA234 (Power supply is 23.4 VDC)
01GA150 (Power supply is 15.0 VDC)
Comments: Voltage value is 3-byte and sends with 0.1 V increment.
I
Read indicated weight
Command : [ADR][I]
Response : [ADR][I][STATUS][SIGN][WEIGHT VALUE]
Example :
Command : 01I
Response : 01IS+000123.4 (Weight is stable and 123.4)
01ID+000123.4 (Weight is dynamic and 123.4)
01I+ (Overload)
Comments:
Indicated weight value (stable or dynamic) is transmitted immediately.
The weight value may be in gross or net.
P
Read the stable weight
Command : [ADR][P]
Response : [ADR][P][STATUS][SIGN][WEIGHT VALUE]
Example :
Command : 01P
Response : 01PS+000123.4 (Weight is stable and 123.4) or
01PN (Could not print)
Comments:
Checks status and it must be stable. Else Nack status is send.
Status can be Stable or Nack.
Q
Load set points
Command : [ADR][Q][SET No][L][SIGN][SP VALUE]
Response : [ADR][Q][STATUS]
Example :
Command : 01Q01L+000123.4
Response : 01QA (123.4 loaded to SP1)
01QN (Could not loaded)
01QX (Decimal point of SP VALUE is mismatch)