FT-111, User Manual Rev. 2.0.0, June 2022 Page 95 of 137
40093
R
2
Gross weight
40127
R/W
2
Dynamic filter
Refer to parameter [332] on page 58
40129
R/W
2
Digital filter
Dec
Description
0
No
1
Very Low
2
Low
3
Medium
4
High
5
Very High
40131
R/W
2
Power on zero
0
Disable
1
± 2%
2
± 2%LK
3
± %10
4
+ %15, - %5
5
± %20
40133
R/W
2
Zeroing Range
0
Disable
1
± 2%
2
± 3%
3
± 20%
4
± 50%
40135
R/W
2
Auto Zero Tracking
0
Disable
1
± 0,3d
2
± 0,5d
3
± 1d
4
± 2d
5
± 3d
40137
R/W
2
Tare
0
No
1
Multi tare
2
Tare only at gross
40139
R/W
2
Stability Detection Range
0
± 0,3d
1
± 0,5d
2
± 1d
3
± 2d
4
± 3d
5
± 4d
6
Disable
40141
R/W
2
Stability Time
Refer to parameter [517] on page 62
40143
R/W
10
Not used
40153
R/W
2
Unit
Dec
Description
0
g
1
kg
2
t
3
lb
4
oz
5
No
40155
R/W
2
Range
0
Single range
1
2 x MR
2
3 x MR
3
2 x MI
4
3 x MI
40157
R/W
2
MAX-1
Refer to parameter [522] on page
40159
R/W
2
Decimal point-1
Dec
Description
0
XXXXOO
1
XXXXXO

Table of Contents