Remote Operation
SCPI Commands and Syntax 5
5-29
The response is:
CURRent An ordinary current range is active.
VOLTage A voltage out of current range is active.
5-49. Output Subsystem Command Details
OUTPut[:STATe](?) <bool>{OFF|ON|0|1}
This command turns the instrument’s output on or off, dependent upon the individual
Voltage and Current channel output settings of each phase.
ON or 1 will set the output on.
OFF or 0 will set the output off.
The query command returns 1 if the output is on, or 0 if the output is off.
OUTPut:ROSCillator[:STATe](?) <bool>{OFF|ON|0|1}
This command turns the instrument’s reference oscillator signal on or off.
ON or 1 will enable the generation of a sample reference signal.
OFF or 0 will disable the generation of a sample reference signal.
The query command returns 1 if reference oscillator is on, or 0 if reference oscillator is off.
OUTPut:SENSe(?) <bool>{OFF|ON|0|1}
This command turns the instrument’s 2-wire or 4-wire sense capability on or off.
ON or 1 will select 4-wire sensing.
OFF or 0 will select 2-wire sensing.
The query command returns 1 if 4-wire sensing is on, or 0 if 2-wire sensing is on.
OUTPut:DEFer[:STATe](?) <bool>{OFF|ON|0|1}
This command sets the deferred or direct operating mode.
When deferred mode is active, all commands that effect the output signal of the master
instrument and phases, are buffered until the instrument receives an apply or undo operation.
At this point, the actual output signal on all phases is updated to reflect the buffered state, or
the buffered state is undone.
ON or 1 will enable deferred mode.
OFF or 0 will disable deferred mode, and return the instrument to direct operation.
The query command returns 1 if deferred mode is on, or 0 if deferred mode is off.
Note
The Instrument will default to direct mode.
OUTPut:DEFer:ACTion <cpd>{APPLy | UNDO}
This command will apply or undo any pending (buffered) command that have been received
when in deferred mode.
APPLy will act upon those commands last received since the last apply/undo.
UNDO will discard any commands received since the last apply.
The command has no query form. Note that it will report a settings conflict if DEFer mode
is not ON.