6100B/6105A
Users Manual
5-44
This selects whether the voltage out of current equivalence factor is applied.
ON or 1 will enable equivalence factor.
OFF or 0 will disable equivalence factor
SOURce:PHASe<x>:CURRent:EQUivalence:FACTor(?) <dnpd>
This command sets the value, in units of Volts per Amp, of the voltage to current
equivalence factor
SOURce:PHASe<x>:CURRent:TERMinal:ROUTe(?) <cpd>{ UPPer | LOWer }
This command selects where the up-to-21A signal is routed for this phase. Used in
conjunction with the the :TERMinal:STATe and :TERMinal:ROUTe commands.
UPPer will select the normal 4 mm terminals
LOWer will select the lower 50 A terminals
Note
This command will return a settings conflict if the 50 A option is not fitted.
SOURce:PHASe<x>:CURRent:BANDwidth(?) [<cpd>{ NORMal | LOW }]
This command is used to select the current channel bandwidth limit.
NORMAL a 6 kHz limit is applied.
LOW a 1.5 kHz limit is applied.
The query command will return the active setting.
Note: the *OPT? command reports whether bandwidth selection is available.
5-69. Harmonics Phenomenon
SOURce:PHASe<x>:CURRent:MHARmonic[:STATe](?) <bool>{OFF|ON|0|1}
This command turns the specified phase’s current channel harmonics phenomena on and off,
toggling it with the Sine mode
ON or 1 will enable harmonics mode, disabling Sine mode.
OFF or 0 will disable harmonics mode, enabling Sine mode.
The query form returns the current state.
SOURce:PHASe<x>:CURRent:MHARmonics:CLEar
This command clears all harmonics, except the fundamental associated with this phase's
Current. It does not have a query form.
SOURce:PHASe<x>:CURRent:MHARmonics:AMPLitude(?) <dnpd>
This command sets the RMS value of the harmonic waveshape. Any harmonics will be
scaled appropriately to keep the waveshape of the composite waveform the same. The query
form returns the RMS value.
SOURce:PHASe<x>:CURRent: MHARmonic:HARMonic<y> <dnpd>,<dnpd>
This command sets the specified phase’s current channel harmonics for harmonic number y
(1 to 100). The parameters specify amplitude (in the presently selected current amplitude
units), and phase angle (in the presently selected phase angle units), respectively.
SOURce:PHASe<x>:CURRent:MHARmonic:HARMonic<y>? [<cpd>{ AMPLitude |
PANGle }]
This query returns the amplitude (in the presently selected current amplitude units), and
phase angle (in the presently selected phase angle units) of the specified harmonic on the
specified phase. Add the appropriate optional parameter to query just one of these values.