Remote Operation
SCPI Commands and Syntax 5
5-35
For reference purposes, note that the following ranges are presently defined for the
6100B/6101B:
Range Lower Limit Upper Limit
15 V range 0.0 V 23 V
30 V range 0.0 V 45 V
60 V range 0.0 V 90 V
120 V range 0.0 V 180 V
230 V range 0.0 V 360 V
560 V range 0.0 V 1008 V
For reference purposes, note that the following ranges are presently defined for the
6105A/6106A:
Range Lower Limit Upper Limit
15 V range 0.0 V 23 V
30 V range 0.0 V 45 V
60 V range 0.0 V 90 V
120 V range 0.0 V 180 V
230 V range 0.0 V 360 V
480 V range 0.0 V 650 V
560 V range 0.0 V 1008 V
SOURce:PHASe<x>:VOLTage:RANGe? [<cpd>{ LOW | HIGH }]
The default version will return the low and high limits of the presently selected range,
comma separated. Add the appropriate optional parameter to query just one of these values.
SOURce:PHASe<x>:VOLTage:AMPLitude?
This query only command is used to find out the specified phase's output amplitude, in
RMS Volts.
The instrument will return the present voltage value. The returned number will be in
standard scientific format (550 V would be returned as 5.5E2).
5-55. DC and Harmonics Phenomenon
SOURce:PHASe<x>:VOLTage:MHARmonics:STATe(?) <bool>{OFF|ON|0|1}
This command turns the specified phase’s voltage channel harmonics phenomena on and off,
toggling it with the sine mode:
• ON or 1 will enable Harmonics mode, disabling sine mode.
• OFF or 0 will disable Harmonics mode, enabling sine mode.