Front-Panel Operation
Edit Mode 4
4-11
Note
6100A setups are not transferrable to 6100B and 6105A and vice versa.
4-17. Set Date and Time
Date and time are adjusted via the Date and Time softkey.
4-18. GPIB Settings
Bus address, Event Status Enable (ESE), Status Register Enable (SRE) and Power On
status Clear (PSC) can be configured with the GPIB Settings pop-up screen.
4-19. Set Factory Defaults
Set Factory Defaults resets all settings including harmonics to the factory default
setting.
4-20. About
The About screen provides information about the current hardware and software
configuration.
4-21. Edit Mode
controls Edit Mode. The next sections explain Edit Mode in detail.
4-22. Direct Mode
In Direct Mode, is lit. All waveform changes take immediate effect.
4-23. Deferred Mode
When is not lit, the Instrument is in Deferred Mode. In this mode, changes made are
stored for later invocation. When in deferred mode, if the output for the channel being
modified is ON, modification to fields that affect the output waveform are notified by an
orange background color.
Note
Operations which are invalid when the output is ON are also invalid when
Deferred Mode is active, even if the output is OFF. For example, you
cannot change range in Deferred Mode even if the output is OFF.
To activate Deferred Mode changes, select the Output Menu softkey Apply All or, if the
output is already ON, press .
The following actions undo all pending changes:
Push the Undo All softkey,
Push
Or
Push (edit mode changes to Direct)
4-24. Changes that are not Deferred
In Deferred Mode, changes to all fields are deferred with the following exceptions:
Line Locking
Change of harmonic edit mode (e.g. Absolute RMS, % of RMS etc)
Power calculation method
Selection of Phase Units (Degrees/Radians)
Selection of two-wire/4 Wire (terminal configuration cannot be changed when the