EasyManua.ls Logo

Fluke 6105A - Harmonics with Side Bands; Phase Jumps

Fluke 6105A
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6100B/6105A
Users Manual
4-28
4-56. Harmonics with Side Bands
The Harmonics with Side Bands Flicker function, see Figure 4-35, allows the input
bandwidth of Flickermeters to be explored. The Fundamental frequency voltage
waveform is modulated by two frequencies simultaneously. Both frequencies are of the
same amplitude.
gdw29.bmp
Figure 4-35. Harmonics with Side Bands
Entering a harmonic number (hn) sets the harmonic frequency (fv) as a multiple of the
fundamental frequency. See Table 4-6.
An interharmonic modulating frequency fi = fv -10 Hz is also applied.
For example:
fundamental frequency = 50 Hz,
hn = 7, fv = 50 * 7 = 350 Hz,
fi = 350 - 10 = 340 Hz.
Table 4-6. Harmonics with Side Bands
120 V 230 V
Fundamental
frequency (Hz)
Starting
frequencies
(Hz)
Modulating
frequency
amplitude (V)
Fundamental
frequency (Hz)
Change to
frequency
(Hz)
Change to
voltage (V)
170 & 180 4.126 140 & 150 3.611
60
50
140 & 150 4.126
50
60
170 & 180 3.611
Flicker meter input bandwidth is the maximum fv frequency at which P
inst,max
is 1.00.
4-57. Phase Jumps
The Phase Jumps Flicker function, see Figure 4-36, causes a series of voltage channel
phase jumps over a 10-minute period. The phase jumps occur at the positive zero
crossing at 1 minute, 3 minutes, 5 minutes, 7 minutes, and 9 minutes after the end of the
settling period. The phase jump direction and size is selected by the operator at the start
of a sequence. Table 4-7 shows the expected P
st
for the different combinations of voltage,
frequency and phase jump size.

Table of Contents

Other manuals for Fluke 6105A

Related product manuals