Introduction
Electrical Specifications 1
1-19
1-43.
Accuracy
Counted/Timed timing accuracy
± ( 10 ppm + 100 ns )
[2]
Gated mode accuracy
± ( 10 ppm + 100 ns )
[2]
Packet mode accuracy (ppm)
[3]
± ( output power (ppm) + 10 ppm + 110,000/Test Duration (secs) )
[2] Accuracy depends on the period between the application of power (pressing the OPER key) and the gate signal
becoming active being greater than 2 seconds.
[3] Specification not valid if “Soft Start” is enabled.
1-44. Test Duration
Maximum test duration 1000 hours
1-45. Power Specifications
The example power specifications below are only valid for rms values greater than 40 % of range for voltage and
current and frequency less than 450 Hz. They are not valid when any of: Flicker, Fluctuating harmonics, Dip/Swell or
Interharmonics are applied to the voltage or current channel of that 6100B.
1-46. 6105A and 6106A Sinusoidal Power Accuracy at 45 Hz to 65 Hz; Power
Factor 1.0 (ppm)
[1]
Power with Current at 90 % of Range Power with Current at 50 % of Range
Voltage at 62 % to 70 % of
Range
Voltage at 7 % to 100 % of
Range
Current
23 V to 90 V 180 V & 360 V
650 V & 1008
V Ranges;
70 % to 75 %
23 V to 90 V 180 V & 360 V
650 V & 1008
V Ranges;
70 % to 75 %
0 A to 2 A 62 64 64 72 74 74
5 A to 50 A 65 66 66 74 75 75
80 A 147 148 148 181 181 181
[1] For Energy specification add 1 ppm
Note: 100 ppm = 0.01 %
1-47. 6105A and 6106A Sinusoidal Power Accuracy at 45 Hz to 65 Hz; Power
Factor 0.5 (ppm)
[1]
Power with Current at 90 % of Range Power with Current at 50 % of Range
Voltage at 62 % to 70 % of
Range
Voltage at 7 % to 100 % of
Range
Current
23 V to 90 V 180 V & 360 V
650 V & 1008 V
Ranges; 70 %
to 75 %
23 V to 90 V 180 V & 360 V
650 V & 1008 V
Ranges; 70 %
to 75 %
0 A to 5 A
93 94 94 100 101 101
10 A to 50 A
95 96 96 102 102 102
80 A
163 163 163 194 194 194
[1] For Energy specification add 1 ppm
Note: 100 ppm = 0.01 %