6100B/6105A
Users Manual
1-20
1-48.
6100B and 6101B Sinusoidal Power Accuracy at 45 Hz to 65 Hz; Power
Factor 1.0 (ppm)
[1]
Power with Current at 90 % of Range Power with Current at 50 % of Range
Current
23 V to 360 V
Ranges; 62 % to
70 % Range
1008 V Range;
740 V to 850 V
23 V to 360 V
Ranges; 62 % to
70 % Range
1008 V Range; 740
V to 850 V
0 A to 2 A 236 239 252 239
5 A to 50 A 236 239 252 239
80 A 322 339 404 417
[1] For Energy specification add 1 ppm
Note: 100 ppm = 0.01 %
1-49. 6100B and 6101B Sinusoidal Power Accuracy at 45 Hz to 65 Hz; Power
Factor 0.5 (ppm)
[1]
Power with Current at 90 % of Range Power with Current at 50 % of Range
Current
23 V to 360 V
Ranges; 62 % to
70 % Range
1008 V Range;
740 V to 850 V
23 V to 360 V
Ranges; 62 % to
70 % Range
1008 V Range; 740
V to 850 V
0 A to 5 A
246 249 262 249
10 A to 50 A
246 249 262 249
80 A
329 346 409 423
[1] For Energy specification add 1 ppm
Note: 100 ppm = 0.01 %
1-50. 6105A and 6106A Sinusoidal Power Accuracy with 20 % THD at Power
Factor 1.0 (ppm)
[1]
Accuracy Depends on Harmonic Order and Amplitudes
Power with Current at 90 % of Range Power with Current at 50 % of Range
Voltage at 62 % to 70 % of Range
Voltage at 7 % to 100 % of
Range
Current
23 V to 90 V
180 V &
360 V
650 V & 1008
V Ranges;
70 % to 75 %
23 V to 90 V 180 V & 360 V
650 V & 1008 V
Ranges; 70 %
to 75 %
0 A to 5 A 97 98 98 103 105 105
10 A to 50 A 98 99 99 105 105 105
80 A 165 165 165 196 196 196
[1] For Energy specification add 1 ppm
Note: 100 ppm = 0.01 %
1-51. 6100B and 6101B Non-sinusoidal Power Accuracy with 20 % THD at Power
Factor 1.0 (ppm)
[1]
Accuracy Depends on Harmonic Order and Amplitudes
V Range Power with Current at 90 % of Range Power with Current at 50 % of Range
Current
23 V to 360 V; 62 %
to 70 % Range
1008 V Range;
740V to 850 V
23 V to 360 V
Ranges; 62 % to
70 % Range
1008 V Range;
740 V to 850 V
0 A to 5 A 242 255 258 255
10 A to 50 A 242 255 258 255
80 A 326 350 408 426
[1] For Energy specification add 1 ppm
Note: 100 ppm = 0.01 %