6100B/6105A
Users Manual
4-24
4-51. 6100B Capability
The implementation of Flicker is separated into two groups- Basic Functions and
Extended Functions. The Basic Functions group allow the depth and frequency of
rectangular and Sine to be chosen for calibration of Flickermeters at the settings in IEC
61000-4-15. See Table 4-2.
Table 4-2. Basic Flicker Function
Function Explanation
Setting range
±30 % of set value within range values (60 %
ΔV/V)
Shape of modulation envelope Rectangular, Square or Sinusoidal
Modulation units Either: Frequency 0.05 Hz to 40 Hz
Or: Changes per
minute
1.0 CPM to 4800 CPM
The Extended Functions provide additional tests with distorted waveforms and
combinations of frequency, amplitude and phase angle changes.
4-52. Setting Up Basic Flicker
Select the Basic Functions softkey from the top level Flicker menu. The Flicker panel
has three sections. The Modulation and Waveform panes set the modulation shape. The
Flicker severity pane shows the P
st
and P
inst
values that the Flickermeter should display.
Flicker parameters can be set within the ranges specified in the previous table. Note that
change rate units can be set to frequency (Hz) or changes per minute (CPM). See Figures
4-30, 4-31, and 4-29.
gdw25.bmp
Figure 4-29. Basic Flicker Softkeys
gdw23.bmp
Figure 4-30. Flicker Menu (Frequency)