Remote Operation
Device Settings After *RST 5
5-69
5-126. Device Settings After *RST
*RST will reset the instrument to a defined condition, stated for each applicable command.
The reset condition is not dependent on past use history of the instrument except as noted below:
*RST does not affect the following:
The selected address of the instrument.
Calibration data that affect specifications.
SRQ mask conditions.
The state of the IEEE 488.1 interface.
The Error Queue.
The Power on Status Clear flag setting.
The contents of:
The Status Byte Register.
The Status Byte Enable Register.
The Standard Event Status Register.
The Standard Event Status Enable Register.
The SCPI Operation Status Register.
The SCPI Operation Status Enable Register.
The SCPI Questionable Status Register.
The SCPI Questionable Status Enable Register.
*RST enforces the following states:
The instrument is returned to “Operation Complete Command Idle State” (OCIS).
The instrument is returned to “Operation Complete Query Idle State” (OQIS.
Settings Related to Common IEEE 488.2 Commands are as detailed in “Common Commands
and Queries:
The “Enable Macro Command” (*EMC) is not used in the instrument.
The “Define Device Trigger Command” (*DDT) is not used in the instrument.
Parallel Poll is not implemented in the instrument.