Remote Operation
Device Settings After *RST 5
5-69
5-126. Device Settings After *RST
*RST will reset the instrument to a defined condition, stated for each applicable command.
The reset condition is not dependent on past use history of the instrument except as noted below:
*RST does not affect the following:
• The selected address of the instrument.
• Calibration data that affect specifications.
• SRQ mask conditions.
• The state of the IEEE 488.1 interface.
• The Error Queue.
• The Power on Status Clear flag setting.
• The contents of:
• The Status Byte Register.
• The Status Byte Enable Register.
• The Standard Event Status Register.
• The Standard Event Status Enable Register.
• The SCPI Operation Status Register.
• The SCPI Operation Status Enable Register.
• The SCPI Questionable Status Register.
• The SCPI Questionable Status Enable Register.
*RST enforces the following states:
• The instrument is returned to “Operation Complete Command Idle State” (OCIS).
• The instrument is returned to “Operation Complete Query Idle State” (OQIS.
Settings Related to Common IEEE 488.2 Commands are as detailed in “Common Commands
and Queries”:
• The “Enable Macro Command” (*EMC) is not used in the instrument.
• The “Define Device Trigger Command” (*DDT) is not used in the instrument.
• Parallel Poll is not implemented in the instrument.