Remote Operation
Common Commands and Queries 5
5-67
5-122. Recall Service Request Enable
This Status Byte data structure conforms to the IEEE 488.2 standard requirements for this
structure.
Figure 5-15. Service Request Enable Query
*SRE? recalls the enable mask for the standard defined events.
Response Decode:
The value returned, when converted to base 2 (binary), identifies the enabled bits that will
generate a service request. The detail is contained in the IEEE 488.2 standard document.
Execution Errors:
None.
Power On and Reset Conditions
The Power On condition depends on the condition stored by the common *PSC command if
0 then it is not cleared. If 1 then the register is cleared. Reset has no effect.
5-123. Read Service Request Register
This Status Byte data structure conforms to the IEEE 488.2 standard requirements for this
structure.
*STB? recalls the service request register for summary bits.
Figure 5-16. Status Byte Query
Response Decode:
The value returned, when converted to base 2 (binary), identifies the summary bits for the
current status of the data structures involved. For the detail definition see the IEEE 488.2
standard document. There is no method of clearing this byte directly. Its condition relies on
the clearing of the overlying status data structure.
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.