Calibration
Calibration Adjustment Process 7
7-15
gdw51.bmp
Figure 7-7. Adjust Instrument Screen
7-26. Determine the 6100B Error
Ensure the measurement equipment is correctly configured; connections are correctly
made and turn the 6100B/6101B output on. After allowing the 6100B/6101B and
measurement equipment time to settle, for each of the components to be adjusted:
1. Note the difference (D) between Target (T) and measured value (M): D=T–M
2. Calculate the required Actual value = T+D (which is the same as 2T–M)
3. Enter the Actual value in the Actual field
7-27. Initiate the Adjustment
Press Accept adjustment.
The 6100B instrument stores the amplitude and phase calibration constants.
Allow the 6100B and measurement equipment time to settle.
If the residual errors are within limits, report the amplitude and phase values to the
calibration certificate.
Otherwise, repeat the calibration adjustment process until the verification measurement
results are within the required tolerance.
7-28. Return Calibration Switch to Normal
Once calibration adjustment is complete, return the calibration enable switch to the
Normal position.