Introduction
Electrical Specifications 1
1-9
1-22.
Voltage Sine Amplitude Specifications
Ranges Frequency Voltage
[5]
6105A & 6106A 1-
Year Accuracy, tcal
[4]
±5 °C ± (ppm of
output + mV)
[1]
6100B & 6101B 1-
Year Accuracy,
tcal
[4]
±5 °C ±
(ppm of output +
mV)
[1]
6100B, 6101B, 6105A &
6106A Open Loop 24-
Hour Stability ± (ppm
of output + mV)
[2][3]
15 V - 17 V 42 0
45 Hz - 65 Hz
1.0 V - 23 V
42 0.2
1.0 V - 23 V
16 Hz - 850 Hz 1.0 V - 23 V
60 0.2
112 1 75 0.8
28 V - 32 V 42 0
45 Hz - 65 Hz
3 V - 45 V
42 0.4
3 V - 45 V
16 Hz - 850 Hz 3 V - 45 V
60 0.4
112 2 75 0.8
56 V - 64 V 42 0
45 Hz - 65 Hz
6.3 V - 90 V
42 0.8
6.3 V - 90 V
16 Hz - 850 Hz 6.3 V - 90 V
60 0.8
112 2.2 75 0.8
110 V - 128 V 44 0
45 Hz - 65 Hz
13 V - 180 V
44 1.6
13 V - 180 V
16 Hz - 850 Hz 13 V - 180 V
60 1.6
112 4.4 75 1.5
215 V - 246 V 44 0
45 Hz - 65 Hz
25 V - 360 V
60 3.2
25 V - 360 V
16 Hz - 850 Hz 25 V - 360 V
61 3.2
112 8.8 75 3
425 V - 490 V 44 0
45 Hz - 65 Hz
46 V - 650 V
60 5.8
46 V – 650 V
16 Hz - 850 Hz 46 V - 650 V
61 5.8
- - 75 6
740 V - 850 V 44 0
45 Hz - 65 Hz
70 V - 1008 V
60 10
70 V - 1008 V
16 Hz - 850 Hz 70 V - 1008 V
61 10
150 26 75 10
[1] Four-wire sense only, for two-wire operation, add an additional voltage = 0.3 Ω x maximum burden current to the accuracy
specification.
[2] For ±1 °C and constant load and connection conditions.
[3] When Flicker, Fluctuating harmonics, Dip/Swell or Interharmonics are applied, Open Loop Stability specification must be added
to the 1-year accuracy specification as described in “Open and Closed Loop Operation”.
[4] tcal = temperature of last calibration.
[5] Output levels less than the range minimum can be set but are not specified.