6100B/6105A
Users Manual
5-66
Execution Errors:
None
Power On and Reset Conditions
No change. This data is saved in non-volatile memory at power off, for use at power on.
5-120. Reset
Figure 5-13. Reset
*RST will reset the instrument to a defined condition, stated for each applicable command
with the command's description, and listed in “Device Settings at Power On”.
The reset condition is not dependent on past use history of the instrument except as noted
below:
*RST does not affect the following:
• The selected address of the instrument.
• Calibration data that affect specifications.
• SRQ mask conditions.
• The state of the IEEE 488.1 interface.
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.
5-121. Service Request Enable
This Status Byte data structure conforms to the IEEE 488.2 standard requirements for this
structure.
Figure 5-14. Service Request Enable
*SRE enables the standard and user defined summary bits in the service request byte, which
will generate a service request.
Nrf is a Decimal Numeric Data Element representing an integer decimal value equivalent to
the Hex value required to enable the appropriate bits in this 8 bit register. The detail
definition is contained in the IEEE 488.2 document.
Note that numbers will be rounded to an integer.
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.