6100B/6105A
Users Manual
1-14
1-28.
Current DC and Harmonic Amplitude Specifications
Range Current
[4]
Frequency
6105A & 6106A
1-Year Accuracy,
tcal
[3]
±5 °C ±(ppm
of output + μA)
6100B & 6101B
1-Year Accuracy,
tcal
[3]
±5 °C ±(ppm
of output + μA)
Open Loop
24-Hour Stability
±(ppm of output
+ μA)
[1][2]
0 A - 0.125 A DC 89 25 139 75 100 11
16 Hz - 850 Hz 61 5 139 6 75 3
0.01 A - 0.25 A
0 A - 0.075 A
850 Hz - 6 kHz 400 5 400 6 150 3
0 A - 0.25 A DC 89 50 139 150 100 22
16 Hz - 850 Hz 61 10 139 12 75 5
0.05 A - 0.5 A
0 A - 0.15 A
850 Hz - 6 kHz 400 10 400 12 150 5
0 A - 0.5 A DC 89 100 139 300 100 45
16 Hz - 850 Hz 61 20 139 24 75 10
0.1 A -1 A
0 A - 0.3 A
850 Hz - 6 kHz 400 20 400 24 150 10
0 A - 1 A DC 89 200 139 600 100 90
16 Hz - 850 Hz 61 40 182 48 75 20
0.2 A - 2 A
0 A - 0.6 A
850 Hz - 6 kHz 400 40 400 48 150 20
0 A - 2.5 A DC 89 500 139 1500 100 225
16 Hz - 850 Hz 61 100 139 120 75 50
0.5 A - 5 A
0 A - 1.5 A
850 Hz - 6 kHz 400 100 400 120 150 50
0 A - 5 A DC 89 1000 191 3000 100 450
16 Hz - 850 Hz 64 200 191 240 75 100
1 A - 10 A
0 A - 3 A
850 Hz - 6 kHz 400 200 400 240 150 100
0 A - 10 A DC 90 2000 191 6000 100 900
16 Hz - 850 Hz 65 400 191 720 75 300
2 A - 21 A
0 A - 6 A
850 Hz - 6 kHz 400 400 400 720 150 300
16 Hz - 850 Hz 69 1000 250 2800 500 750
5 A - 50 A 0 A - 15 A
850 Hz - 3 kHz 400 1000 400 2800 750 1200
16 Hz - 850 Hz 112 2000 265 2800 500 1200
8 A - 80 A 0 A - 24 A
850 Hz - 3 kHz 400 2000 400 2800 750 1200
[1] For ±1 °C and constant load and connection conditions.
[2] When Flicker, Fluctuating harmonics, Dip/Swell or Interharmonics are applied, “Open loop” stability specification must be added to
the 1-year accuracy specification as described in “Open and Closed Loop Operation”.
[3] tcal = temperature of last calibration.
[4] These specifications are only applicable if the combined voltage rms output is greater than the range minimum. If the combined
output is below the range minimum the output is not specified.
[5] The maximum value for a single harmonic (2nd to 100th) below 2850 Hz is 30 % of range. See “Amplitude/Frequency Limits” for
profile above 2850 Hz.