Two-Wire / Four-Wire Voltage Settings 3
3-13
gdw13.bmp
Figure 3-9. Waveform Menu, Current Output
3-14. Turn the Output Off
To turn the output off, push . The on indications disappear.
The Set up exercise is complete.
3-15. Two-Wire / Four-Wire Voltage Settings
The Instrument voltage output can be configured for either two-wire or four-wire sensing.
The default mode is two-wire sense.
XWWarning
When the Instrument voltage channel is enabled and operating,
full voltage output may be present at the VOLTAGE SENSE
terminals even when they are not externally connected. To
avoid electrical shock, personal injury, use caution when using
the instrument under these conditions.
The 6100B is provided with a four to two-wire lead set for voltage. Irrespective of two-
wire / four-wire configuration, the four plugs at the 6100B end of the lead should always
be plugged in when the lead is in use. The sense wires will always be at the same
potential as the output wires.
For measurement accuracy it is important that the two-wire / four-wire setting is correct.
In particular, setting four-wire sensing but only connecting two wires will cause voltage
errors in the order of 1.2 V. If you are not familiar with four-wire voltage sensing please
read the description below.