Introduction
Electrical Specifications 1
1-17
1-35.
Current to Voltage Phase Specifications
For Voltage from the Current terminals, use the 0.25 A to 21 A phase specification.
For All Voltage Ranges
(23 V - 1008 V)
6105A /6106A Voltage and
Current Components >40 % of
Range
6100B / 6101B Voltage and
Current Components >40 %
of Range
Voltage or Current
Component 0.5 % - 40 %
of Range
[5]
Current
Range
Frequency
1-
Y
ear Accuracy,
tcal
[4]
±5 °C
[1][2]
Open Loop
24-Hour
Stability
[2][3]
1-Year
Accuracy,
tcal
[4]
±5 °C
[1][2]
Open Loop
24-Hour
Stability
[2][3]
1-Year
Accuracy,
tcal ±5 °C
[1][2]
Open Loop
24-Hour
Stability
[2][3]
45 Hz - 65 Hz
0.0023 ° 0.0002 ° 0.003 ° 0.0002 ° 0.010 ° 0.001 °
16 Hz - 69 Hz
0.003 ° 0.0002 ° 0.003 ° 0.0002 ° 0.010 ° 0.001 °
69 Hz - 180 Hz
0.007 ° 0.0002 ° 0.009 ° 0.0002 ° 0.017 ° 0.002 °
180 Hz - 450 Hz
0.018 ° 0.0005 ° 0.023 ° 0.0005 ° 0.050 ° 0.005 °
450 Hz - 850 Hz
0.033 ° 0.0008 ° 0.043 ° 0.0008 ° 0.070 ° 0.007 °
850 Hz - 3 kHz
0.115 ° 0.0010 ° 0.150 ° 0.0010 ° 0.200 ° 0.020 °
0.25 A - 21 A
3 kHz - 6 kHz
0.230 ° 0.0010 ° 0.300 ° 0.0010 ° 0.450 ° 0.045 °
45 Hz - 65 Hz
0.0023 ° 0.0002 ° 0.004 ° 0.0003 ° 0.010 ° 0.001 °
16 Hz - 69 Hz
0.003 ° 0.0003 ° 0.004 ° 0.0003 ° 0.010 ° 0.001 °
69 Hz - 180 Hz
0.007 ° 0.0003 ° 0.012 ° 0.0003 ° 0.017 ° 0.002 °
180 Hz - 450 Hz
0.018 ° 0.0005 ° 0.030 ° 0.0005 ° 0.050 ° 0.005 °
450 Hz - 850 Hz
0.033 ° 0.0010 ° 0.050 ° 0.0010 ° 0.070 ° 0.007 °
850 Hz - 3 kHz
0.115 ° 0.0015 ° 0.200 ° 0.0015 ° 0.200 ° 0.020 °
21 A - 50 A
3 kHz - 6 kHz
0.230 ° 0.0025 ° 0.300 ° 0.0025 ° 0.450 ° 0.045 °
45 Hz - 65 Hz
0.003 ° 0.0002 ° 0.004 ° 0.0005 ° 0.010 ° 0.001 °
16 Hz - 69 Hz
0.003 ° 0.0005 ° 0.004 ° 0.0005 ° 0.016 ° 0.002 °
69 Hz - 180 Hz
0.008 ° 0.0005 ° 0.012 ° 0.0005 ° 0.028 ° 0.003 °
180 Hz - 450 Hz
0.025 ° 0.0010 ° 0.030 ° 0.0010 ° 0.080 ° 0.008 °
450 Hz - 850 Hz
0.050 ° 0.0015 ° 0.050 ° 0.0015 ° 0.100 ° 0.010 °
20 A - 80 A
850 Hz - 3 kHz
0.250 ° 0.0025 ° 0.200 ° 0.0025 ° 0.300 ° 0.030 °
[1] Current phase angle errors are relative to the voltage channel of the same phase. For example, L2 current is relative to L2 voltage.
[2] Phase angle contribution to power accuracy varies with set phase angle see “Power Specifications” below.
[3] For constant load and connection conditions.
[4] tcal = temperature of last calibration.
[5] Phase performance at less than 0.5 % of full range degrades as output components approach the resolution limit of the digital
feedback system.