6100B/6105A
Users Manual
5-60
IPWH Impulses per Watt-hour
IPKWH Impulses per killoWatt-hour
IPKMH Impulses per MegaWatt-hour
WHIP Watt-hour per impulse
KWHIP killoWatt-hour per impulse
KMHIP MegaWatt-hour per impulse
Note that there is no short form of these parameters.
5-103. Reference Source
ENERgy:REFerence:SOURce(?) { CH6|SUM456|MEAN456|MEAN56|MMUT|MAIN }
This command specifies the pulse source(s) used to define the reference source.
The source(s) can be:
CH6 Pulse stream from (independent) channel 6.
SUM456 Sum of channels 4 through 6.
MEAN456 Arithmetic mean of channels 4 through 6.
MEAN56 Arithmetic mean of channels 5 and 6.
MMUT Mean of the active MUT sources (typically used in “drift monitor” scenarios).
MAIN Equivalent pulse count of the power outputs (value calculated from the
“OUTPut:CONStant” meter constant value and active power settings).
Note
If the requested source channel is already in use as part of a MUT source
definition, an error will be reported.
5-104. Reference Pull-up
ENERgy:REFerence:PULLup <cpd> { R150|R1000 }
This command specifies which internal pull-up resistance value to select for the reference
source:
R150 150 Ohm.
R1000 1 k Ohm.
5-105. Output Tree
The output commands configure the pulse out channel. This channel generates a pulse
stream that is proportional to the active power outputs.
Pulses will be generated whenever the energy timer/counter option has been selected, and
the output is on.
5-106. Output Meter Constant
ENERgy:OUTPut:CONStant(?) <dnpd>
This command specifies the meter constant (as impulses per unit energy) used by the output
channel.