6100B/6105A
Users Manual
5-52
The Query will return the updated time at the moment the query was accepted as
HH,MM,SS.
SYSTem:VERSion?
The query only command returns an <Nr2> formatted numeric value corresponding to the
SCPI version number for which the 6100B complies. At the time of writing, this will be
1999.0.
5-76. Unit Subsystem Command Details
UNIT:ANGLe(?) <cpd>{DEGrees|RADians}
This command sets the units to be used to express all instances of Phase Angle.
DEG will set the phase angle units to be Degrees.
RAD will set the phase angle units to be Radians.
The query command will return DEG if the units are set to Degrees, or RAD if the units are
set to Radians.
UNIT:MHARmonics:CURRurrent(?) <cpd>{PRMS|PFUN|DBF|ABS }
This command selects the specified harmonics amplitude units for current.
PRMS will set the units to be Percentage of RMS Current amplitude.
PFUN will set the units to be Percentage of Fundamental amplitude.
DBF will set the units to be dB down from Fundamental amplitude.
ABS will set the units to be Absolute value.
The query command will return PRMS if the units are set to “Percentage of RMS Current”,
etc.
UNIT:MHARmonics:VOLTage(?) <cpd>{PRMS|PFUN|DBF|ABS}
This command selects the specified harmonics amplitude units for voltage.
PRMS will set the units to be Percentage of RMS Voltage amplitude.
PFUN will set the units to be Percentage of Fundamental amplitude.
DBF will set the units to be dB down from Fundamental amplitude.
ABS will set the units to be Absolute value.
The query command will return PRMS if the units are set to percentage of RMS voltage, etc.
UNIT:DIP:TIME(?) <cpd>{SEConds|CYCLes}
This command selects the units used for time, when specifying dip parameters.
SEC will set the Dip time units to be seconds.
CYCL will set the Dip time units to be cycles.
The query command will return SEC if the Dip time units are set to seconds etc.
UNIT:FLICker:CURRent:FREQuency(?) <cpd> {HZ|CPM}
This command selects the units for change rate when specifying flicker parameters for
current:
HZ will set the change rate to Hertz.
CPM will set the change rate to Changes Per Minute
The query command will return HZ or CPM.
UNIT:FLICker:VOLTage:FREQuency(?) <cpd> {HZ|CPM}
This command selects the units for change rate when specifying flicker parameters for
voltage: