Remote Operation
Common Commands and Queries 5
5-63
5-114. *IDN? (Instrument Identification)
This command conforms to the IEEE 488.2 standard requirements.
Figure 5-7. Instrument Identification
*IDN? will recall the instrument’s manufacturer, model number, serial number and
firmware level.
Response Format:
Character position
Fluke,6100B,XXXXXXXXXXXX,X.XX or Fluke,6105A,XXXXXXXXXXXX,X.XX
Where:
The data contained in the response consists of four comma-separated fields, the last two of
which are instrument-dependent. The data element type is defined in the IEEE 488.2
standard specification.
Response Decode:
The data contained in the four fields is organized as follows:
First field - manufacturer.
Second field - model.
Third field - serial number.
Fourth field - firmware level (will possibly vary from one instrument to another).
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.
5-115. Operation Complete
This command conforms to the IEEE 488.2 standard requirements.
Figure 5-8. Operation Complete
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.
*OPC is a synchronization command which will generate an operation complete message in
the standard Event Status Register when all pending operations are complete.