6100B/6105A
Users Manual
1-10
1-23.
Voltage DC and Harmonic Amplitude Specifications
Range Output
[4][5]
Frequency
6105A & 6106A
1-Year Accuracy,
tcal
[4]
±5 °C ±(ppm
of output + mV)
[1][6]
6100B & 6101B
1-Year Accuracy, tcal
[4]
±5 °C ±(ppm of
output + mV)
[1][6]
Open Loop 24-
Hour Stability ±
(ppm of output +
mV)
[2][3]
0 V - 11.5 V DC 91 2 122 5 75 1.8
16 Hz - 850 Hz 58 1 122 1 75 0.8
1.0 V - 23 V
0 V - 6.9 V
850 Hz - 6 kHz
451 1 512 1 150 0.8
0 V - 22.5 V DC 91 4 122 10 75 3.3
16 Hz - 850 Hz 58 2 122 2 75 0.8
3 V - 45 V
0 V - 13.5 V
850 Hz - 6 kHz
451 2 512 2 150 0.8
0 V - 45 V DC 91 8 122 24 75 8
16 Hz - 850 Hz 60 2.2 122 2.2 75 0.8
6.3 V - 90 V
0 V - 27 V
850 Hz - 6 kHz
451 2.2 512 2.2 150 0.8
0 V - 90 V DC 91 16 122 50 75 15
16 Hz - 850 Hz 60 4.4 122 4.4 75 1.5
13 V - 180 V
0 V - 54 V
850 Hz - 6 kHz
451 4.4 512 4.4 150 1.5
0 V - 180 V DC 91 32 122 100 75 30
16 Hz - 850 Hz 60 12 122 12 75 3
25 V - 360 V
0 V - 108 V
850 Hz - 6 kHz
451 12 512 12 150 3
0 V - 325 V DC 92 60 - - 75 65
16 Hz - 850 Hz 61 22 - - 75 6
46 V - 650 V
0 V - 195 V
850 Hz - 6 kHz
451 22 - - 150 6
0 V - 504 V DC 92 100 166 300 75 100
16 Hz - 850 Hz 61 33 166 33 75 10
70 V -
1008 V
0 V - 302 V
850 Hz - 6 kHz
451 33 524 33 150 10
[1] Four-wire sense only, for two-wire operation, add an additional voltage = 0.3 Ω x maximum burden current to the accuracy
specification.
[2] For ±1 °C and constant load and connection conditions.
[3] When Flicker, Fluctuating harmonics, Dip/Swell or Interharmonics are applied, “Open loop” stability specification must be added to
the 1-year accuracy specification as described in “Open and Closed Loop Operation”.
[4] These specifications are only applicable if the combined voltage rms output is greater than the range minimum. If the combined
output is below the range minimum the output is not specified.
[5] The maximum value for a single harmonic (2nd to 100th) below 2850 Hz is 30 % of range. See “Amplitude/Frequency Limits” for
profile above 2850 Hz.
[6] tcal = temperature of last calibration.