Remote Operation
SCPI Commands and Syntax 5
5-19
5-45. SCPI Commands and Syntax
5-46. SCPI Command Summary
Keyword Parameter Form Notes
CALibration
:SECure
:PASSword <spd> Used to enter calibration mode: Requires
calibration password.
:EXIT Used to exit calibration mode.
:PHASe<x> <x> is phase (1 to 4): 1 is master phase.
:VOLTage
:RANGe <dnpd>, <dnpd> Calibration range:
<dnpd> = Low limit, High limit.
:RANGe? [<cpd> {LOW | HIGH}] Calibration range query.
:ACTual <dnpd>,<dnpd> <dnpd> = Amplitude, Angle.
Note: Angle relative to master phase.
:FREQuency? Query only of Actual frequency.
:TARGet <dnpd>[,<dnpd>,<dnpd>,<dnpd>,<dnpd>] <dnpd> = Target point.
Or,
<dnpd> =
Point,
Fund Freq,
Harmonic,
Amplitude,
Angle.
Note1: Angle relative to master phase.
Note2: Second form is only required when
changing target point.
:TRIGger?
:STORe
:DUMP? Dump all stores for active range:
Point,(<target data>,<actual data>)
<target data> =
Fund, Harm, Ampl, Angle
<actual data> =
Freq, Ampl, Angle