Remote Operation
SCPI Commands and Syntax 5
5-47
AMPLitude | FREQuency}]
The default version of this query returns all of the settings of the specified inter-harmonic,
comma separated. Add the appropriate optional parameter to query just one of these values.
5-72. Dip Phenomenon
SOURce:PHASe<x>:CURRent:DIP:STATe(?) <bool>{OFF|ON|0|1}
This command turns the specified Phase’s Current channel Dip phenomena on and off.
ON or 1 will set the specified Dip to be applied.
OFF or 0 will set the specified Dip to be removed.
The query command will return 1 if Dip is applied, or 0 if Dip is inactive.
SOURce:PHASe<x>:CURRent:DIP:ENVelope <dnpd>,<dnpd>,<dnpd>,<dnpd>,<dnpd>
This command sets the specified Phase’s current channel Dip parameters:
1
st
dnpd - Change To value (expressed as a percentage of total RMS voltage).
2
nd
dnpd - Ramp In period (expressed in Seconds or Cycles).
3
rd
dnpd - Duration (expressed in Seconds or Cycles).
4
th
dnpd - Ramp Out period (expressed in Seconds or Cycles).
5
th
dnpd - End Delay period (expressed in Seconds or Cycles).
SOURce:PHASe<x>:CURRent:DIP:ENVelope? [<cpd>{CHANge | RIN | DURation |
ROUT | EDELay}]
The default version of this query returns the Dip Envelope settings for the specified phase's
current channel. Add the appropriate optional parameter to query just one of these values:
CHANge
Change To value, expressed as a percentage of the total RMS Voltage
RIN
Ramp In period, expressed in Seconds or Cycles depending on the Dip Units setting
DURation
Duration, expressed in Seconds or Cycles depending on the Dip Units setting
ROUT
Ramp Out period, expressed in Seconds or Cycles depending on Dip Units setting
EDELay
End Delay period, expressed in Seconds or Cycles
SOURce:PHASe<x>:CURRent:DIP:TRIGger:INPut(?) <cpd>{ FREE | EONE |
EREPeat}
This command sets and queries the trigger mode used to determine the event that starts the
dip or swell.
FREE is used for free running dips/swells.
EONE is used to produce one dips/swell triggered from an external source.
EREPeat is used to produce continuous dips/swells triggered from an external source.
SOURce:PHASe<x>:CURRent:DIP:TRIGger:HOLDoff (?) <cpd>{PHASe|DELay},<dnpd>
This command selects sets and queries the hold-off before the dip/swell starts following a
trigger:
PHASe The hold-off is an angle following the trigger point. In this case the delay ,<dnpd>, has units
of degrees or radians.
DELay The hold-off is a time. In this case the delay ,<dnpd>, has units of seconds or cycles.