6100B/6105A
Users Manual
1-16
1-33.
DC and Harmonic Amplitude Specifications
Range Output
[4][5]
Frequency
6105A & 6106A 1-Year
Accuracy, tcal
[4]
±5 °C ±
(ppm of output + μV)
[5]
6100B & 6101B
1-Year Accuracy, tcal
[4]
±5 °C ± (ppm of output
+μV)
[5]
Open Loop Stability
± (ppm of output +
μV) per Hour
[2][3]
0 V - 0.125 V DC 91 35 200 35 100 15
16 Hz - 850 Hz 82 10 200 10 60 15
0.05 V -
0.25 V
0 V - 0.075 V
850 Hz - 6 kHz 400 30 1000 30 150 15
0 V - 0.75 V DC 93 210 200 210 100 75
16 Hz - 850 Hz 66 35 200 50 50 25
0.15 V -
1.5 V
0 V - 0.45 V
850 Hz - 6 kHz 400 50 1000 50 150 25
0 V - 5 V DC 93 1000 200 1000 100 450
16 Hz - 850 Hz 65 200 200 200 50 150
1 V - 10 V
0 V - 3 V
850 Hz - 6 kHz 400 300 1000 300 150 150
[1] tcal = temperature of last calibration.
[2] For ±1 °C and constant load and connection conditions.
[3] When Flicker, Fluctuating harmonics, Dip/Swell or Interharmonics are applied, “Open loop” stability specification must be added
to the 1-year accuracy specification as described in “Open and Closed Loop Operation”.
[4] These specifications are only applicable if the combined voltage rms output is greater than the range minimum. If the combined
output is below the range minimum the output is not specified.
[5] The maximum value for a single harmonic (2nd to 100th) below 2850 Hz is 30 % of range. See “Amplitude/Frequency Limits” for
profile above 2850 Hz.
1-34. Voltage from Current Terminals, Distortion and Noise
Range and Frequency
Maximum Harmonic Distortion
[1]
Either
Non-harmonic Noise Floor
(relative to full range)
the largest of or the largest of 16 Hz - 4 MHz
Full
Range
Frequency
dB Volts % Setting % Range dB %
16 Hz - 850 Hz -80
2.5 μV
0.010 0.001 -50 0.316
0.25 V
850 Hz - 6 kHz -60
25 μV
0.100 0.01 -50 0.316
16 Hz - 850 Hz -80
15 μV
0.010 0.001 -60 0.100
1.5 V
850 Hz - 6 kHz -60
150 μV
0.100 0.01 -60 0.100
16 Hz - 850 Hz -80
100 μV
0.010 0.001 -60 0.100
10 V
850 Hz - 6 kHz -60 1 mV 0.100 0.01 -60 0.100
[1] dB harmonic distortion increases linearly between 50 Hz and 6 kHz.