5-124. Test Operations — Full Selftest
This query conforms to the IEEE 488.2 standard requirements.
Figure 5-17. Test Query
*TST? executes a Full selftest. A response is generated after the test is completed.
N. B. Operational selftest is valid only at temperatures: 23 °C ±10 °C.
Response Decode:
The value returned identifies pass or failure of the operational selftest:
ZERO indicates operational selftest complete with no errors detected.
Non zero indicates operational selftest has failed. The number itself represents the
number of test failures.
The failure codes can be found only by re-running the self-test manually.
Execution Errors:
Operational selftest is not permitted when calibration is successfully enabled.
Power On and Reset Conditions
Not applicable.
5-125. Wait
This command conforms to the IEEE 488.2 standard requirements.
Figure 5-18. Wait
*WAI prevents the instrument from executing any further commands or queries until the No
Pending Operations Flag is set true. This is a mandatory command for IEEE-488.2 but has
little relevance to this instrument as there are no parallel processes requiring Pending
Operation Flags.
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.