Remote Operation
SCPI Commands and Syntax 5
5-41
5-61. Extended Flicker State
[SOURce]:PHASe<x>:VOLTage:EFLicker[:STATe](?) <bool> { ON | OFF | 0 | 1 }
Enable or disable the currently selected extended flicker signal configuration (for the passed
phase x). If the output is on, the new configuration will be applied immediately to the output.
5-62. Configure Signal
[SOURce]:PHASe<x>:VOLTage:EFLicker:CONFigure(?) <cpd> { PF | MZ | HS | PJ | RV }
Select the extended flicker signal function (for the passed phase x).
Note
The instruments state will not change until :EFLicker:STATe is set to ON (or
1 ). If the extended flicker function is changed, :EFLicker:STATe will
automatically be set back to OFF or (0).
The extended functions are:
PF - Flicker signal with periodic frequency changes (section 4.3.3 of IEC 601000-4).
MZ - Distorted voltage with multiple zero crossings (section 4.3.43 of IEC 601000-4).
HS - Harmonics with sideband (section 4.3.5 of IEC 601000-4).
PJ - Phase jumps (section 4.3.6 of IEC 601000-4).
RV - Rectangular voltage changes with duty cycle (section 4.3.7 of IEC 601000-4).
The default value is PF (periodic frequency changes).
5-63. Select Sideband Harmonic
[SOURce]:PHASe<x>:VOLTage:EFLicker:HSIDeband:HARMonic(?) <dnpd>
Select the distorting harmonic to use when generating the harmonics with side-band test
signal.
Value range: 3-99.
Value default: 3.
Note
This value is only applied when HSIDeband is selected using
the :EFLicker:CONFigure command.
5-64. Select Phase Jump Angle
[SOURce]:PHASe<x>:VOLTage:EFLicker:PJUMp:ANGLe(?) < dnpd >
Select the phase angle to use with the phase jump sequence test: only values of +/- 30.0
degrees or +/- 45.0 degrees will be accepted.
The default value is +30.0.
Note
This value is only applied when PJUMp is selected using
the :EFLicker:CONFigure command.
5-65. Select Phase Jump Settle Period
[SOURce]:PHASe<x>:VOLTage:EFLicker:PJUMp:SPERiod(?) <cpd> { OFF | S5 |
S10 | M1 | M5 | M10 }
Select the settle period to use before starting the phase jump sequence. The output is on
during this period: