Remote Operation
Common Commands and Queries 5
5-61
5-107. Output Meter Constant Units
ENERgy:OUTPut:CONStant:UNIT(?){ IPWH|IPKWH|
IPMWH|WHPI|KWHPI|MWHPI }
This command specifies the meter constant units (as impulses per unit energy) used by
channels considered to be MUT sources.
IPWH Impulses per Watt-hour
IPKWH Impulses per killoWatt-hour
IPKMH Impulses per MegaWatt-hour
WHIP Watt-hour per impulse
KWHIP killoWatt-hour per impulse
KMHIP MegaWatt-hour per impulse
Note that there is no short form of these parameters.
5-108. Output Pull-up
ENERgy:OUTPut:PULLup[:STATe] <bool> { ON|OFF|0|1 }
This command specifies whether the output channel should switch in an internal pulled up
resistor.
5-109. Common Commands and Queries
5-110. Clear Status
This measurement event status data structure conforms to the IEEE 488.2 standard
requirements for this structure.
Figure 5-3. Clear Status
*CLS clears all the event registers and queues except the output queue.
The output queue and MAV bit will be cleared if *CLS immediately follows a “Program
Message Terminator”; refer to the IEEE 488.2 standard document.
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.
5-111. Event Status Enable
This event status data structure conforms to the IEEE 488.2 standard requirements for this
structure.
Figure 5-4. Event Status Enable
*ESE enables the standard defined event bits, which will generate a summary message in