6100B/6105A
Users Manual
1-6
• Data transfer via external storage
• Operator maintenance, including calibration
• Glossary of Terms
1-8. Specifications
The following sections give the Instrument’s general and detailed specifications.
1-9. General Specifications
1-10. Input Power
Voltage
100 V - 240 V with up to ±10 % fluctuations
Transient overvoltages Impulse withstand (overvoltage) category II of IEC 60364-4-443
Frequency 47 Hz - 63 Hz
Max. Consumption 1000 VA max from 100 - 130 V, 1250 VA max from 130 V - 260 V
1-11. Dimensions
6100B, 6101B, 6105A, and 6106A With 50 A or 80 A Options
Height 233 mm (9.17 inches) 324 mm (12.8 inches)
Height (without feet) 219 mm (8.6 inches) 310 mm (12.2 inches)
Width 432 mm (17 inches) 432 mm (17 inches)
Depth 630 mm (24.8 inches) 630 mm (24.8 inches)
Weight 23 kg (51 lb) 30 kg (66 lb)
1-12. Environment
Operating temperature
5 °C - 35 °C
Calibration temperature (tcal) range
16 °C - 30 °C
Storage temperature
0 °C - 50 °C
Transit temperature
-20 °C - 60 °C <100 hours
Warm up time 1 hour
Safe operating max. relative humidity (non-condensing)
<80 % 5 °C - 31 °C ramping linearly down to 50 % at
35 °C
Storage max. relative humidity (non-condensing)
<95 % 0 °C - 50 °C
Operating altitude 0 m - 2,000 m
Non-operating altitude 0 m - 12,000 m
Shock MIL-PRF-28800F class 3
Vibration MIL-PRF-28800F class 3
Enclosure MIL-PRF-28800F class 3
1-13. Safety
• Meets CAN/CSA-C22.2 No 61010.1-04, UL Std. No. 61010-1 (2
nd
Edition), ISA-82.02.01
Reference Standard: IEC 61010-1:2001
• Indoor use only, pollution degree 2; installation category II
• CE marked and CSA listed
1-14. EMC
EN61326-1:2006, CISPR 11 Class A, FCC rules part 15, sub-part B, class A (Class A equipment is suitable for use in
establishments other than domestic, and those directly connected to a low voltage power supply network which supplies
buildings used for domestic purposes).