6100B/6105A
Users Manual
6-4
6-4. Changing the User Password
Navigate to the Waveform Menu using . If necessary push until the top-level
softkey menu is displayed. See Figure 6-1.
gdw32.bmp
Figure 6-1. Waveform Menu Top Level Softkeys
Select Support Functions, Diagnostic Tools, and enter the user password. The
default password when the 6100B is first shipped is “12321”.
Select Change Password to display the “change the calibration password...” pop-up
menu. Enter the existing password, the new password and the new password again. Push
to change the password (or to cancel the operation).
6-5. Accessing the Fuse
The power fuse is accessible from the rear panel. See Figure 6-2.
XWWarning
To avoid electrical shock or personal injury, ensure that the
Instrument is switched off at the rear-mounted on/off switch
and disconnected by removing the line power cord from the
power input socket before attempting to access the power fuse,
To access the fuse, proceed as follows:
1. Disconnect line power.
2. Using a standard screwdriver, turn the fuse holder counterclockwise until the cap and
fuse are disengaged.
Always replace with the approved fuse shown in Table 6-1.
Table 6-1. Approved Replacement Fuses
Description Part Number Rating
Fluke part number and description: 1998159 T15AH 250 V 32 mm
Fuse manufacturer and part number: Bussmann MDA-15