EasyManua.ls Logo

Fluke 6105A - Multi-Phase Operation; Voltage to Voltage Phase Specifications; Energy Specifications; Pulse Inputs

Fluke 6105A
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6100B/6105A
Users Manual
1-18
1-36.
Multi-Phase Operation
1-37. Voltage to Voltage Phase Specifications
For All Voltage
Ranges
(23 V - 1008 V)
6105A /6106A Voltage and
Current Components >40 %
of Range
6100B / 6101B Voltage and
Current Components >40 %
of Range
Voltage or Current
Component 0.5 % - 40 % of
Range
[5]
Frequency
1-Year
Accuracy,
tcal
[4]
±5
°C
[1][2]
Stability per
hour
[2][3]
1-Year
Accuracy,
tcal
[4]
±5
°C
[1][2]
Stability per
hour
[2][3]
1-Year
Accuracy,
tcal ±5 °C
[1][2]
Stability per
hour
[2][3]
16 Hz - 69 Hz
0.005 ° 0.0002 ° 0.005 ° 0.0002 ° 0.010 ° 0.001 °
69 Hz - 180 Hz
0.007 ° 0.0002 ° 0.007 ° 0.0002 ° 0.017 ° 0.002 °
180 Hz - 450 Hz
0.025 ° 0.0005 ° 0.025 ° 0.0005 ° 0.050 ° 0.005 °
450 Hz - 850 Hz
0.043 ° 0.0008 ° 0.050 ° 0.0008 ° 0.070 ° 0.007 °
850 Hz - 3 kHz
0.150 ° 0.0010 ° 0.170 ° 0.0010 ° 0.200 ° 0.020 °
3 kHz - 6 kHz
0.300 ° 0.0010 ° 0.350 ° 0.0015 ° 0.450 ° 0.045 °
[1] Current phase angle errors are relative to the voltage channel of the same phase. For example, L2 current is relative to L2
voltage.
[2] Phase angle contribution to power accuracy varies with set phase angle see “Power Specifications” below.
[3] For constant load and connection conditions.
[4] tcal = temperature of last calibration.
[5] Phase performance at less than 0.5 % of full range degrades as output components approach the resolution limit of the
digital feedback system.
1-38. Energy Specifications
1-39. Pulse Inputs
Max frequency 5 MHz (100 Hz for debounced inputs)
Min pulse width 100 ns
Max counts per channel 232-1 (4,294,967,295)
1-40. Pulse and Gate Inputs
Input Low level max 1 V
Input High level min 3 V
135 Ω and 940 Ω to 4.5 V nominal
Internal pull-up values
(Approximately equivalent to 150 Ω/1k Ω to 5 V nominal)
Max input voltage 28 V (clamped @ 30 V approximately)
[1]
Min input voltage 0 V (clamped @ -0.5 V approximately)
[1]
1-41. Pulse Output
Drive
Open-collector with optional internal 470 Ω pull-up
Frequency range 1 mHz – 5 MHz
Frequency accuracy
± ( 10 ppm + 100 nHz )
External pull-up voltage 30 V MAX (clamped)
[1]
Sink current 150 mA MAX
1-42. Gate Output
Drive Open-drain
Internal Pull-up As Gate-Input
External pull-up voltage 30 V MAX (clamped)
[1]
Sink current 1 A MAX
[1] Input/Output protection: 30 V / -0.5V (approximately) clamped, up to 120 mA per signal or 300 mA maximum total all
signals.

Table of Contents

Other manuals for Fluke 6105A