6100B/6105A
Users Manual
5-58
Note: If one of the source parameters has been changed after using this command a valid
default will be set (typically “MAIN”); so it is recommend that the “MUT:TEST:PSOUrce”
command is sent after any “MUT:SOURce” or “REFerence:SOURce” commands, to
prevent unexpected side effects.
The source can be:
CH1 to CH6 Raw pulse streams from individual channels.
SUM456 Sum of channels 4 through 5.
MEAN456 Arithmetic mean of channels 4 through 6.
MEAN56 Arithmetic mean of channels 5 and 6.
EMUT On Every MUT channel. For example, on selecting a test
duration of 100 pulses, at least 100 pulses would have to be
counted on all selected MUTs before the test sequence was
considered complete.
MAIN Equivalent pulse count of the power outputs (value calculated
from the “OUTPut:CONStant” meter constant value and active
power settings).
5-92. ENERgy: ABORt
This command allows an active test sequence to be aborted. Its action is similar to hitting
the ‘STBY’ key or sending ‘:OUTPut OFF’, with one exception: when the ‘Maintain
Voltage’ output feature has been enabled (ENERgy:ENERgy:MVOLTage), the voltage
channels will remain on (even though the test sequence has been reset to the beginning
again). If ':OUTPut OFF' is used to force a test sequence to stop, all channels will turn off
regardless of whether ‘Maintain Voltage’ has been set.”
5-93. MUT Tree
The MUT commands configure a MUT source. A MUT source is typically made up from
pulse streams from channels 1 to 3, though channels 4 to 6 can also be used when treated as
independent sources (i.e. for “drift monitoring”).
5-94. MUT Meter Constant
ENERgy:MUT:CONStant(?) <dnpd>
This command specifies the meter constant (as impulses per unit energy) used by channels
considered to be MUT sources.
5-95. MUT Meter Constant Units
ENERgy:MUT:CONStant:UNIT(?){ IPWH|IPKWH|
IPMWH|WHPI|KWHPI|MWHPI }
This command specifies the meter constant units (as impulses per unit energy) used by
channels considered to be MUT sources.
IPWH Impulses per Watt-hour
IPKWH Impulses per killoWatt-hour
IPKMH Impulses per MegaWatt-hour
WHIP Watt-hour per impulse
KWHIP killoWatt-hour per impulse
KMHIP MegaWatt-hour per impulse