5-76
5-132. Energy Option Action on Receiving *RST
The following default settings are used on receiving a *RST command.
Setting Value Following *RST
[:SOURce]
:ENERgy
:MODE(?) Timed/counted mode.
:UNIT(?) Real (Wh).
:OGATe(?) Output gate Off.
Gate type level.
Gate active low.
Internal pull-up 1k Ohm.
:IGATe(?) Gate type level.
Gate active low.
Internal pull-up 1k Ohm.
:MVOLtage(?) 0 (all modes)
:PRESentation(?) Accumulated Energy, % Error.
:RESults(?) n/a
:WUP
:DURation(?) SEC,100
:PSOurce(?) MAIN
:TEST
:DURation(?) SEC,600
:PSOurce(?) MAIN
:MUT
:DEBounce
[:STATe](?) 0
:CONStant(?) 1.0e5 i/Wh
:UNIT IPWH
:SOURce(?) CH1
:PULLup(?) 1 k Ohm.
:REFerence
:DEBounce
[:STATe](?) 0
:CONStant(?) 1.0e5 i/Wh
:UNIT IPWH
:SOURce(?) MAIN
:PULLup(?) 1 k Ohm.
:OUTPut
:CONStant(?) 1.0e5 i/Wh
:UNIT IPWH
:PULLup[:STATe](?) 0