6100B/6105A
Users Manual
5-70
5-127. Device Settings at POWER ON
5-128. General
Active Mode: The instrument powers up in “manual” mode.
Device I/D (Serial Number). Factory serial number preserved.
Status Reporting Conditions:
Status Byte Register. Depends on state of *PSC.
Status Byte Enable Register. Depends on state of *PSC.
Event Status Register. Depends on state of *PSC.
Event Status Enable Register. Depends on state of *PSC.
Operation Status Event Register. Depends on state of *PSC.
Operation Status Enable Register. Depends on state of *PSC.
Questionable Status Event Register. Depends on state of *PSC.
Questionable Status Enable Register. Depends on state of *PSC.
Error Queue. Empty until first error is detected.
5-129. Power-On Settings Related to Common IEEE 488.2 Commands
Program Coding Condition
*CLS Not applicable
*ESE Nrf Not applicable
*ESE? Response depends on state of *PSC
*ESR? Response depends on state of *PSC
*IDN? Not applicable
*OPC Not applicable
*OPC? Not applicable
*PSC 0/ 1 Not applicable
*PSC? No change. This data is saved at power off for use at power on.
*PUD Data area remains unchanged
*PUD? Data area remains unchanged
*RST Not applicable
*SRE Nrf Not applicable
*SRE? Response depends on state of *PSC
*STB? Response depends on state of *PSC
*TST? Not applicable
*WAI Not applicable