6100B/6105A
Users Manual
4-26
4-54. Periodic Frequency Changes
The Periodic Frequency Changes Flicker function, see Figure 4-33, provides a fixed
pattern of changes every 4 seconds. Frequency is stepped ±0.25 Hz either side of the
fundamental frequency while voltage steps by up to 1.2 V depending on voltage and
fundamental frequency settings. It should be noted that in a multiphase system the ±0.25
Hz frequency changes will occur on every phase. The voltage changes will occur only on
the selected voltage channel. See Table 4-3.
gdw27.bmp
Figure 4-33. Combined Frequency and Voltage Changes
Table 4-3. Periodic Frequency Changes
120 V 230 V
Fundamental
Frequency
(Hz)
Change to
Frequency
(Hz)
Change to
Voltage (V)
Fundamental
Frequency
(Hz)
Change to
Frequency
(Hz)
Change to
Voltage (V)
59.75 120.000 49.75 230.000
60
60.25 119.266
50
50.25 228.812
49.75 120.000 59.75 230.000
50
50.25 119.270
60
60.25 228.805
The observed P
inst.max
should be 1.00.