6100B/6105A
Users Manual
7-16
7-29. Verification
Verify adjustments using the same equipment against the product specification in
Chapter 1.
7-30. Calibration Adjustment Points
The following tables list the calibration adjustment points.
Table 7-8. Voltage Adjustment Points
Range
(Volts)
Fundamental
Frequency
(Hz)
Harmonic
Number
Frequency
(Hz)
Point
Output Level
Setting
(Volts)
Output Phase
Angle Setting
(Degrees)
0 0
DC
offset
0.000 0
1 57 Gain 16.000 0
57
15 855 Gain 16.000 0
23
63 95 5985 Gain 16.000 0
0 0
DC
offset
0.000 0
1 57 Gain 30.000 0
57
15 855 Gain 30.000 0
45
63 95 5985 Gain 30.000 0
0 0
DC
offset
0.000 0
1 57 Gain 60.000 0
57
15 855 Gain 60.000 0
90
63 95 5985 Gain 60.000 0
0 0
DC
offset
0.000 0
1 57 Gain 120.000 0
57
15 855 Gain 120.000 0
180
63 95 5985 Gain 120.000 0
0 0
DC
offset
0.000 0
1 57 Gain 240.000 0
57
15 855 Gain 240.000 0
360
63 95 5985 Gain 240.000 0
0 0
DC
offset
0.000 0
1 57 Gain 460.000 0
57
15 855 Gain 460.000 0
650
(6105A
and
6106A
only)
63 95 5985 Gain 400.000 0
0 0
DC
offset
0.000 0
1 57 Gain 800.000 0
57
15 855 Gain 800.000 0
1008
63 95 5985 Gain 400.000 0